Chromatographic conditions for single column setup

Parameter Agilent 1290 Infinity I/II LC
Analytical column Agilent Zorbax Eclipse Plus C18, 1.8 µm, 100 x 2.1 mm column, p/n 959758-902
Inline filter Agilent 1290 Infinity II inline filter, 0.3µm, p/n 5067-6189
Column temperature 45 °C
Injection volume 1 µL (equivalent to 0.1 µL plasma)
Autosampler temperature 20 °C
Needle wash Wash vial, 3 s, 1:1 butanol/methanol
*Alternatively, flush port, 10 s, 1:1 mobile phase A/B can be used
Mobile phase A = 10 mM ammonium formate, 5µM Agilent Deactivator Additive (5191-3940) in 5:3:2 water/acetonitrile/2-propanol
B = 10 mM ammonium formate in 1:9:90 water/acetonitrile/2-propanol
Flow rate 0.4 mL/min
Gradient program Time %B
0 15
2.5 50
2.6 57
9 70
9.1 93
11 96
11.1 100
12 100
12.2 15
16 15
Stop time 16 min
Post time None

Solvent composition