Chromatographic conditions for dual column setup

Parameter Agilent 1290 Infinity I/II LC
Analytical columns 2X Agilent Zorbax Eclipse Plus C18, 1.8 µm, 100 x 2.1 mm column, p/n 959758-902 
Inline filters 2X Agilent 1290 Infinity II inline filter, 0.3µm, p/n 5067-6189
Column temperature 45 °C
Injection volume 1 µL (equivalent to 0.1 µL plasma)
Autosampler temperature 20 °C
Needle wash Wash vial, 3 s, 1:1 butanol/methanol
*Alternatively, flush port, 10 s, 1:1 mobile phase A/B can be used
Mobile phase A = 10 mM ammonium formate, 5µM Agilent Deactivator Additive (5191-3940) in 5:3:2 water/acetonitrile/2-propanol
B = 10 mM ammonium formate in 1:9:90 water/acetonitrile/2-propanol
Flow rates 0.4 mL/min
Gradient program Pump A (G7120A) Pump B (G7120A)
Time %B Time %B
0 15 0 15
2.5 50 1 15
2.6 57 2 100
9 70 6 100
9.1 93 7 15
11 96 12.9 15
11.1 100    
12 100    
12.2 15    
12.9 15    
Stop time 13 min
Post time None
MCT Valve Position Use current, Increase valve position after run