| Analytical columns |
2X Agilent Zorbax Eclipse Plus C18, 1.8 µm, 100 x 2.1 mm column, p/n 959758-902 |
| Inline filters |
2X Agilent 1290 Infinity II inline filter, 0.3µm, p/n 5067-6189 |
| Column temperature |
45 °C |
| Injection volume |
1 µL (equivalent to 0.1 µL plasma) |
| Autosampler temperature |
20 °C |
| Needle wash |
Wash vial, 3 s, 1:1 butanol/methanol |
| *Alternatively, flush port, 10 s, 1:1 mobile phase A/B can be used |
| Mobile phase |
A = 10 mM ammonium formate, 5µM Agilent Deactivator Additive (5191-3940) in 5:3:2 water/acetonitrile/2-propanol |
| B = 10 mM ammonium formate in 1:9:90 water/acetonitrile/2-propanol |
| Flow rates |
0.4 mL/min |
| Gradient program |
Pump A (G7120A) |
Pump B (G7120A) |
| Time |
%B |
Time |
%B |
| 0 |
15 |
0 |
15 |
| 2.5 |
50 |
1 |
15 |
| 2.6 |
57 |
2 |
100 |
| 9 |
70 |
6 |
100 |
| 9.1 |
93 |
7 |
15 |
| 11 |
96 |
12.9 |
15 |
| 11.1 |
100 |
|
|
| 12 |
100 |
|
|
| 12.2 |
15 |
|
|
| 12.9 |
15 |
|
|
| Stop time |
13 min |
| Post time |
None |
| MCT Valve Position |
Use current, Increase valve position after run |