Chromatographic conditions

Parameter Agilent 1290 Infinity II LC
Analytical column Agilent Zorbax Eclipse Plus C18, 1.8 µm, 100 x 2.1 mm column, p/n 959758-902
Inline filter InfinityLab Quick Change Inline Filters, 0.2µm, p/n 5067-1603
Column temperature 45 °C
Injection volume 1-5 µL (sample dependent)
Autosampler temperature 17 °C
Needle wash Wash vial, 3 s, 1:1 butanol/methanol
*Alternatively, flush port, 10 s, 1:1 mobile phase A/B can be used
Mobile phase A = 10 mM ammonium formate, 5µM Agilent Deactivator Additive (5191-3940) in 5:3:2 water/acetonitrile/2-propanol
B = 10 mM ammonium formate in 1:9:90 water/acetonitrile/2-propanol
Flow rate 0.4 mL/min
Gradient program Time %B
0 15
2.5 50
2.6 57
12 70
12.5 80
15.4 90
15.5 100
16.5 100
17 15
20 15
Stop time 20 min
Post time None
Gradient program 0-1 minutes To waste
1-18.5 minutes To MS
18.5-20 minutes To waste

Solvent composition